Apical Extrusion of Debris in Flat-oval Root Canals after Using Different Instrumentation Systems

A. L. Kirchhoff, L. F. Fariniuk, and I. Mello

Journal of Endodontics. 2015 February; 41:237–241.

 

INTRODUCTION: The purpose of this study was to evaluate the amount of apically extruded debris in flat-oval root canal systems during cleaning and shaping with different instrumentation systems.

METHODS: Seventy-two mandibular incisors were randomly assigned to 4 groups (n = 18) according to the instrumentation system to be used: the ProTaper Next rotary system (PTN; Dentsply Tulsa Dental, Tulsa, OK), the WaveOne reciprocating system (WO, Dentsply Tulsa Dental), the Twisted File Adaptive (TFA; SybronEndo, Orange, CA) rotary/reciprocating system, and the Self-Adjusting File (SAF) system (ReDent-Nova, Ra'anana, Israel). Debris extruded during the instrumentation was collected and dried in preweighed Eppendorf tubes. Three consecutive weights were obtained for each tube, and the average was calculated. The data were analyzed using the t test and analysis of variance (Games-Howell test).

RESULTS: The SAF produced significantly more debris compared with the other systems (P < .05). No significant difference was noted amongst the PTN, WO, and TFA (P > .05).

CONCLUSIONS: Regardless of the instrumentation system used, apical extrusion of debris was exhibited at some level, regardless of the instrumentation system used. The SAF was associated with the highest amount of debris extrusion compared with PTN, WO, and TFA.

 

Der vollständige Text des Artikels: ResearchGate

Kirchhoff extrusion jendod joe

  

View abstract on PubMed. PMID: 25447504
Melden Sie sich für die neuesten Nachrichten an

Abonnieren Sie unseren Newsletter

Invalid Input

ReDent Nova GmbH & Co. Kg.

Am Borsigturm 70
13507 Berlin, Germany